Abstract
A catalytic immuno-reactor for the determination of human serum albumin (HSA) was constructed by using immobilized antibody and an amperometric detector. A sandwich assay with hemin-labeled antibody to catalyze the decomposition of hydrogen peroxide was used, the catalytic activity of the hemin-antibody conjugate being determined by measuring the decrease in hydrogen peroxide concentration. The reaction of hemin-labeled antibody with antigen was complete within 30 min and the current decrease was correlated with the HSA concentration. The relative standard deviation was about 9% at an HSA concentration of 1 mg ml-1.
Original language | English |
---|---|
Pages (from-to) | 283-287 |
Number of pages | 5 |
Journal | Analytica Chimica Acta |
Volume | 156 |
Issue number | C |
DOIs | |
Publication status | Published - 1984 |
Externally published | Yes |
ASJC Scopus subject areas
- Biochemistry
- Analytical Chemistry
- Spectroscopy
- Environmental Chemistry