T1 - A performance test of a new high-surface-quality and high-sensitivity CR-39 plastic nuclear track detector – TechnoTrak
AU - Kodaira, S.
AU - Morishige, K.
AU - Kawashima, H.
AU - Kitamura, H.
AU - Kurano, M.
AU - Hasebe, Nobuyuki
AU - Koguchi, Y.
AU - Shinozaki, W.
AU - Ogura, K.
PY - 2016/9/15
Y1 - 2016/9/15
N2 - We have studied the performance of a newly-commercialized CR-39 plastic nuclear track detector (PNTD), “TechnoTrak”, in energetic heavy ion measurements. The advantages of TechnoTrak are derived from its use of a purified CR-39 monomer to improve surface quality combined with an antioxidant to improve sensitivity to low-linear-energy-transfer (LET) particles. We irradiated these detectors with various heavy ions (from protons to krypton) with various energies (30–500 MeV/u) at the heavy ion accelerator facilities in the National Institute of Radiological Sciences (NIRS). The surface roughness after chemical etching was improved to be 59% of that of the conventional high-sensitivity CR-39 detector (HARZLAS/TD-1). The detectable dynamic range of LET was found to be 3.5–600 keV/μm. The LET and charge resolutions for three ions tested ranged from 5.1% to 1.5% and 0.14 to 0.22 c.u. (charge unit), respectively, in the LET range of 17–230 keV/μm, which represents an improvement over conventional products (HARZLAS/TD-1 and BARYOTRAK). A correction factor for the angular dependence was determined for correcting the LET spectrum in an isotropic radiation field. We have demonstrated the potential of TechnoTrak, with its two key features of high surface quality and high sensitivity to low-LET particles, to improve automatic analysis protocols in radiation dosimetry and various other radiological applications.
AB - We have studied the performance of a newly-commercialized CR-39 plastic nuclear track detector (PNTD), “TechnoTrak”, in energetic heavy ion measurements. The advantages of TechnoTrak are derived from its use of a purified CR-39 monomer to improve surface quality combined with an antioxidant to improve sensitivity to low-linear-energy-transfer (LET) particles. We irradiated these detectors with various heavy ions (from protons to krypton) with various energies (30–500 MeV/u) at the heavy ion accelerator facilities in the National Institute of Radiological Sciences (NIRS). The surface roughness after chemical etching was improved to be 59% of that of the conventional high-sensitivity CR-39 detector (HARZLAS/TD-1). The detectable dynamic range of LET was found to be 3.5–600 keV/μm. The LET and charge resolutions for three ions tested ranged from 5.1% to 1.5% and 0.14 to 0.22 c.u. (charge unit), respectively, in the LET range of 17–230 keV/μm, which represents an improvement over conventional products (HARZLAS/TD-1 and BARYOTRAK). A correction factor for the angular dependence was determined for correcting the LET spectrum in an isotropic radiation field. We have demonstrated the potential of TechnoTrak, with its two key features of high surface quality and high sensitivity to low-LET particles, to improve automatic analysis protocols in radiation dosimetry and various other radiological applications.
KW - CR-39
KW - Dosimetry
KW - Heavy ions
KW - Nuclear fragmentation
KW - Nuclear track detectors
KW - Space radiation
UR - http://www.scopus.com/inward/record.url?scp=84978402728&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84978402728&partnerID=8YFLogxK
U2 - 10.1016/j.nimb.2016.07.002
DO - 10.1016/j.nimb.2016.07.002
M3 - Article
AN - SCOPUS:84978402728
SN - 0168-583X
VL - 383
SP - 129
EP - 135
JO - Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms
JF - Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms
ER -