Abstract
We describe a novel automated flow immunoassay system for quantification of anti-double-stranded (ds) DNA autoimmune antibodies in the serum of patients suffering from systemic lupus erythematosus, dsDNA (360 bp) was covalently coupled with alkaline phosphatase (ALP) to form a novel analytical reagent (ALP-DNA). After immunoreaction, antibody-antigen complexes between ALP-DNA and anti-dsDNA monoclonal antibody were separated from unreacted ALP- DNA by an ion-exchange column on the basis of the difference in isoelectric point. Antibody-antigen complexes were subsequently quantified by luminescence following addition of 3-(2'-spiroadamantane)-4-methoxy-4-(3- phosphoryloxy)phenyl-1,2-dioxetane. The assay yielded a linear relationship between signal and concentration of anti-dsDNA monoclonal antibody in the range of 0-300 μg/mL. This simple technique permits the assay of anti-dsDNA autoimmune antibodies within 25 min. The ion-exchange column was simply regenerated by occasional elution with eluent (20 mM N-methylpiperazine, pH 5.5) supplemented with 0.5 M NaCl, to remove unreacted ALP-DNA.
Original language | English |
---|---|
Pages (from-to) | 1298-1302 |
Number of pages | 5 |
Journal | Analytical Chemistry |
Volume | 71 |
Issue number | 7 |
DOIs | |
Publication status | Published - 1999 Apr 1 |
Externally published | Yes |
ASJC Scopus subject areas
- Analytical Chemistry