Construction of an electron analyzer with a wide acceptance solid angle

Daisuke Sakai*, Daisuke Miura, Takeshi Nakajima, Tsuyoshi Ishikawa, Chuhei Oshima, Hideo Iwai, Yoshimasa Nihei

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

5 Citations (Scopus)

Abstract

We have developed a new electron energy analyzer with a large solid angle of 0.14π, which is comparable to that of cylindrical mirror analyzer. Typical energy resolution was ΔE/E0 ∼ 0.016 for the aperture of 1 mm and central radius of 100 mm, and typical angular resolution was less than 0.5°.

Original languageEnglish
JournalSurface Science
Volume600
Issue number4
DOIs
Publication statusPublished - 2006 Feb 15

Keywords

  • Auger electron spectroscopy (AES)
  • Electron energy loss spectroscopy (EELS)
  • Electron spectroscopy
  • Low energy electron diffraction (LEED)

ASJC Scopus subject areas

  • Physical and Theoretical Chemistry
  • Condensed Matter Physics
  • Surfaces and Interfaces

Fingerprint

Dive into the research topics of 'Construction of an electron analyzer with a wide acceptance solid angle'. Together they form a unique fingerprint.

Cite this