Abstract
In order to increase the ion-detection efficiency of pressure measurement using nonresonant laser ionization method, a Channeltron of aperture 19 mm was used in the ion-detection head and pressure measurement was carried out with H2. The detection efficiency was increased about 1.5 times compared to a previous result. The volume of the ionization region estimated from the present measurement, 5.8 × 10-4 mm3, is close to that derived from the direct observation of the ion distribution in the ionization region, 6.4 × 10-4 mm3.
Original language | English |
---|---|
Pages (from-to) | 727-729 |
Number of pages | 3 |
Journal | Vacuum |
Volume | 48 |
Issue number | 7-9 |
DOIs | |
Publication status | Published - 1997 Sept |
Externally published | Yes |
ASJC Scopus subject areas
- Instrumentation
- Condensed Matter Physics
- Surfaces, Coatings and Films