TY - JOUR
T1 - In vivo deployment of mechanically adaptive nanocomposites for intracortical microelectrodes
AU - Harris, J. P.
AU - Hess, A. E.
AU - Rowan, S. J.
AU - Weder, C.
AU - Zorman, C. A.
AU - Tyler, D. J.
AU - Capadona, J. R.
PY - 2011/8
Y1 - 2011/8
N2 - We recently introduced a series of stimuli-responsive, mechanically adaptive polymer nanocomposites. Here, we report the first application of these bio-inspired materials as substrates for intracortical microelectrodes. Our hypothesis is that the ideal electrode should be initially stiff to facilitate minimal trauma during insertion into the cortex, yet become mechanically compliant to match the stiffness of the brain tissue and minimize forces exerted on the tissue, attenuating inflammation. Microprobes created from mechanically reinforced nanocomposites demonstrated a significant advantage compared to model microprobes composed of neat polymer only. The nanocomposite microprobes exhibit a higher storage modulus (E′ = ∼5 GPa) than the neat polymer microprobes (E′ = ∼2 GPa) and can sustain higher loads (∼12 mN), facilitating penetration through the pia mater and insertion into the cerebral cortex of a rat. In contrast, the neat polymer microprobes mechanically failed under lower loads (∼7 mN) before they were capable of insertion into cortical tissue. Further, we demonstrated the material's ability to morph while in the rat cortex to more closely match the mechanical properties of the cortical tissue. Nanocomposite microprobes that were implanted into the rat cortex for up to eight weeks demonstrated increased cell density at the microelectrode-tissue interface and a lack of tissue necrosis or excessive gliosis. This body of work introduces our nanocomposite-based microprobes as adaptive substrates for intracortical microelectrodes and potentially for other biomedical applications.
AB - We recently introduced a series of stimuli-responsive, mechanically adaptive polymer nanocomposites. Here, we report the first application of these bio-inspired materials as substrates for intracortical microelectrodes. Our hypothesis is that the ideal electrode should be initially stiff to facilitate minimal trauma during insertion into the cortex, yet become mechanically compliant to match the stiffness of the brain tissue and minimize forces exerted on the tissue, attenuating inflammation. Microprobes created from mechanically reinforced nanocomposites demonstrated a significant advantage compared to model microprobes composed of neat polymer only. The nanocomposite microprobes exhibit a higher storage modulus (E′ = ∼5 GPa) than the neat polymer microprobes (E′ = ∼2 GPa) and can sustain higher loads (∼12 mN), facilitating penetration through the pia mater and insertion into the cerebral cortex of a rat. In contrast, the neat polymer microprobes mechanically failed under lower loads (∼7 mN) before they were capable of insertion into cortical tissue. Further, we demonstrated the material's ability to morph while in the rat cortex to more closely match the mechanical properties of the cortical tissue. Nanocomposite microprobes that were implanted into the rat cortex for up to eight weeks demonstrated increased cell density at the microelectrode-tissue interface and a lack of tissue necrosis or excessive gliosis. This body of work introduces our nanocomposite-based microprobes as adaptive substrates for intracortical microelectrodes and potentially for other biomedical applications.
UR - http://www.scopus.com/inward/record.url?scp=79960771078&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=79960771078&partnerID=8YFLogxK
U2 - 10.1088/1741-2560/8/4/046010
DO - 10.1088/1741-2560/8/4/046010
M3 - Article
C2 - 21654037
AN - SCOPUS:79960771078
SN - 1741-2560
VL - 8
JO - Journal of Neural Engineering
JF - Journal of Neural Engineering
IS - 4
M1 - 046010
ER -