TY - JOUR
T1 - Monte Carlo study of a small field of view YAlO3:Ce pinhole camera for imaging with Rb-82 by detection of bremsstrahlung X-rays
AU - Nakanishi, K.
AU - Yamamoto, S.
N1 - Publisher Copyright:
© 2021 IOP Publishing Ltd and Sissa Medialab.
PY - 2021/8
Y1 - 2021/8
N2 - One of the applications of single-photon emission computed tomography (SPECT) and positron emission tomography (PET) is myocardial imaging. Myocardial perfusion imaging with PET (MPI-PET) is gradually becoming an alternative to MPI-SPECT due to its higher image quality. Although Rb-82 is the most common tracer for MPI-PET, Rb-82 emits high-energy positrons with a long stopping range, resulting in blurring of the spatial resolution of the PET image. Due to the limitations of spatial resolution, imaging of Rb-82 in mice has not been reported. In this study, we propose a new method to achieve higher resolution imaging of Rb-82 in small animals than possible with PET imaging by detecting bremsstrahlung X-rays emitted by the positrons, and we validated the feasibility of this method using Monte Carlo simulation. We simulated a small field of view (FOV) pinhole X-ray camera based on a thin YAlO3:Ce (YAP(Ce)) plate and analyzed the basic performance of the simulated camera for bremsstrahlung X-rays. The spatial resolution of a 0.5 mm-thick YAP(Ce) plate-based camera with a 1.0 mm pinhole collimator was 2.6 mm full width at half maximum (FWHM) at a distance of 17.5 mm from the surface of the collimator. Furthermore, we simulated imaging of a mouse heart phantom filled with Rb-82 of 67 MBq per milliliter. We observed the shape of the phantom in the image for a 10 - 45 keV energy window in a simulated measurement time of 4 minutes. We conclude that imaging of high-energy positron emitters at a higher resolution than by PET imaging is possible through detection of the bremsstrahlung X-rays emitted from the positrons.
AB - One of the applications of single-photon emission computed tomography (SPECT) and positron emission tomography (PET) is myocardial imaging. Myocardial perfusion imaging with PET (MPI-PET) is gradually becoming an alternative to MPI-SPECT due to its higher image quality. Although Rb-82 is the most common tracer for MPI-PET, Rb-82 emits high-energy positrons with a long stopping range, resulting in blurring of the spatial resolution of the PET image. Due to the limitations of spatial resolution, imaging of Rb-82 in mice has not been reported. In this study, we propose a new method to achieve higher resolution imaging of Rb-82 in small animals than possible with PET imaging by detecting bremsstrahlung X-rays emitted by the positrons, and we validated the feasibility of this method using Monte Carlo simulation. We simulated a small field of view (FOV) pinhole X-ray camera based on a thin YAlO3:Ce (YAP(Ce)) plate and analyzed the basic performance of the simulated camera for bremsstrahlung X-rays. The spatial resolution of a 0.5 mm-thick YAP(Ce) plate-based camera with a 1.0 mm pinhole collimator was 2.6 mm full width at half maximum (FWHM) at a distance of 17.5 mm from the surface of the collimator. Furthermore, we simulated imaging of a mouse heart phantom filled with Rb-82 of 67 MBq per milliliter. We observed the shape of the phantom in the image for a 10 - 45 keV energy window in a simulated measurement time of 4 minutes. We conclude that imaging of high-energy positron emitters at a higher resolution than by PET imaging is possible through detection of the bremsstrahlung X-rays emitted from the positrons.
KW - coronary CT angiography (CTA)
KW - Gamma camera
KW - PET PET/_ACTY
KW - Scintigraphy and whole-body imaging
KW - SPECT
KW - X-ray detectors
UR - http://www.scopus.com/inward/record.url?scp=85112511984&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85112511984&partnerID=8YFLogxK
U2 - 10.1088/1748-0221/16/08/T08005
DO - 10.1088/1748-0221/16/08/T08005
M3 - Article
AN - SCOPUS:85112511984
SN - 1748-0221
VL - 16
JO - Journal of Instrumentation
JF - Journal of Instrumentation
IS - 8
M1 - T08005
ER -