抄録
A particle collector that allows high time resolution monitoring of particulate matter was fabricated and coupled with an ion chromatograph for online analysis of particulate ions in the atmosphere. The system was applied to the atmospheric observation for over a month in Tokushima, Japan. The average particulate anion concentrations were 6.42 nmol m-3 for Cl-, 18.8 nmol m-3 for NO-3 , and 22.1 nmol m-3 for SO24-. The atmospheric particle collection efficiency was more than 98.4%, and the continuous observations were successfully achieved without problems.
寄稿の翻訳タイトル | Online Collector for Water-Soluble Atmospheric Particulate Matter Compatible with a Parallel Plate Wet Denuder |
---|---|
本文言語 | Japanese |
ページ(範囲) | 273-278 |
ページ数 | 6 |
ジャーナル | Earozoru Kenkyu |
巻 | 36 |
号 | 4 |
DOI | |
出版ステータス | Published - 2021 |
Keywords
- Air Pollutant
- Continuous Analysis
- Ion Chromatograph
- Online Analysis
- Particle Collector
ASJC Scopus subject areas
- 環境科学(その他)
- 汚染
- 環境化学
- 材料科学(その他)