ウエットデニューダーに対応した水溶性大気粒子状物質のオンライン捕集器

Kaho Maeda, Haruka Tanaka, Riona Wada, Hiroshi Okochi, Kei Toda, Hideji Tanaka, Masaki Takeuchi*

*この研究の対応する著者

研究成果: Article査読

抄録

A particle collector that allows high time resolution monitoring of particulate matter was fabricated and coupled with an ion chromatograph for online analysis of particulate ions in the atmosphere. The system was applied to the atmospheric observation for over a month in Tokushima, Japan. The average particulate anion concentrations were 6.42 nmol m-3 for Cl-, 18.8 nmol m-3 for NO-3 , and 22.1 nmol m-3 for SO24-. The atmospheric particle collection efficiency was more than 98.4%, and the continuous observations were successfully achieved without problems.

寄稿の翻訳タイトルOnline Collector for Water-Soluble Atmospheric Particulate Matter Compatible with a Parallel Plate Wet Denuder
本文言語Japanese
ページ(範囲)273-278
ページ数6
ジャーナルEarozoru Kenkyu
36
4
DOI
出版ステータスPublished - 2021

Keywords

  • Air Pollutant
  • Continuous Analysis
  • Ion Chromatograph
  • Online Analysis
  • Particle Collector

ASJC Scopus subject areas

  • 環境科学(その他)
  • 汚染
  • 環境化学
  • 材料科学(その他)

フィンガープリント

「ウエットデニューダーに対応した水溶性大気粒子状物質のオンライン捕集器」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル