A BGO detector unit for a stationary high resolution positron emission tomograph

Seiichi Yamamoto*, Shuichi Miura, Hidehiro Iida, Iwao Kanno

*この研究の対応する著者

研究成果: Article査読

11 被引用数 (Scopus)

抄録

A new encoding scheme was developed that is applicable to bismuth germanate (BGO) detector units for a high resolution positron emission tomograph. The detector unit is composed of eight equal-size, 3 mm wide BGO crystals coupled to a dual-module square photomultiplier tube (PMT) through a pair of light guides. The light guides divide the scintillation light between the two modules of the PMT, with a ratio dependent on the crystal of origin. Mean detector pair spatial resolution as measured by the point spread function was 3.9 mm full width at half maximum (FWHM) and reconstructed image spatial resolution was 4.8 mm FWHM at the center of the field of view. Timing resolution between two detector units was 6.0 ns FWHM. Energy resolution was 24% FWHM for 511 keV gamma rays. The finest spots of the Derenzo phantom were clearly resolved.

本文言語English
ページ(範囲)851-855
ページ数5
ジャーナルJournal of Computer Assisted Tomography
10
5
DOI
出版ステータスPublished - 1986
外部発表はい

ASJC Scopus subject areas

  • 放射線学、核医学およびイメージング