A linear-array freehand 3-D endoscopic ultrasound

Kazuki Sumiyama*, Naoki Suzuki, Hisao Tajiri

*この研究の対応する著者

研究成果: Article査読

11 被引用数 (Scopus)

抄録

Recognition of the clinical importance of linear-array endoscopic ultrasound (EUS) has increased. In this study, we developed a linear-array 3-D EUS, a miniature position sensor attached to the tip of the echoendoscope used in freehand scanning. To evaluate the geometrical accuracy of the 3-D reconstruction of the system, the diameter of a sphere-shaped phantom (38 mm) was determined by five examiners and five measurers. Measured size of the sphere was 39.03 ± 1.29 mm, with variance between examiners and measurers, and interaction of examiners with measurers was not significant. In animal and clinical studies, the system facilitated anatomical interpretation of the EUS images, especially in the pancreatobiliary area and vascular images. We concluded that this system is both accurate and reproducible, and may resolve difficulties in linear-array EUS.

本文言語English
ページ(範囲)1001-1006
ページ数6
ジャーナルUltrasound in Medicine and Biology
29
7
DOI
出版ステータスPublished - 2003 7月 1
外部発表はい

ASJC Scopus subject areas

  • 生物理学
  • 放射線技術および超音波技術
  • 音響学および超音波学

フィンガープリント

「A linear-array freehand 3-D endoscopic ultrasound」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル