A new microbial electrode using immobilized Clostridium butyricum was prepared for biochemical oxygen demand (BOD) estimation of wastewaters. The current of the electrode was decreased with time until a steady state was reached. The steady state current was in all cases attained within 30-40 min at 37°C, and the maximum current output was obtained at 37°C and pH between 6.2 and 7.0. A linear relationship was obtained between the steady state current and BOD. The steady state current values were reproducible within ±7% of the relative error. The BOD of industrial wastewaters can be estimated by using the microbial electrode. Relative error of the BOD estimation of industrial wastewaters was within ± 10%. The current output of the microbial electrode was almost constant for 30 days.
本文言語 | English |
---|---|
ページ(範囲) | 97-104 |
ページ数 | 8 |
ジャーナル | Journal of Solid-Phase Biochemistry |
巻 | 2 |
号 | 2 |
DOI | |
出版ステータス | Published - 1977 6月 |
外部発表 | はい |
ASJC Scopus subject areas
- 応用微生物学とバイオテクノロジー
- バイオテクノロジー
- 生化学