A stopped-flow apparatus with a push–pull coupler for the study of fast reactions in solution under high pressure

Koji Ishihara, Yuichi Kondo, Mitsuo Koizumi

研究成果: Letter査読

10 被引用数 (Scopus)

抄録

A new high-pressure stopped-flow apparatus equipped with a push–pull coupler has been developed which enables one to monitor fast reactions in various organic solvents as well as in strongly acidic media under pressure up to 200 MPa. Sample solutions come into contact only with chemically inert materials: Daiflon, quartz, and a Teflon-coated O-ring. The reactant-syringe and receiver-syringe pistons are driven at the same time with the aid of a push–pull coupler and an actuator. The dead time of this apparatus is ∼2 ms.

本文言語English
ページ(範囲)244-245
ページ数2
ジャーナルReview of Scientific Instruments
70
1
DOI
出版ステータスPublished - 1999 1月