Assessment of oxygen transfer in hyperbaric dual mode liquid-type oxygenator for ECMO

T. Tsuji*, K. Tanishita, M. Takahashi, T. Togawa

*この研究の対応する著者

研究成果: Conference contribution

抄録

A hyperbaric membrane oxygenator made of ethylene and vinyl alcohol copolymer was developed for extracorporal membrane oxygenation (ECMO). Either gas carrier solution oxygenated with hyperbaric (up to 1460 mm Hg) oxygen gas was perfused or hyperbaric O2 itself was ventilated extracapillarily while saline solution or bovine blood was perfused intracapillarily in in vitro experiments. Sherwood number increased as transmembrane pressure (TMP) increased. Membrane resistance decreased as the TMP increased. The membrane resistance was estimated to be zero at TMP of 1060 mm Hg. This oxygenator can work either as an oxygenator and dialyzer in liquid-liquid mode or as an oxygenator in gas-liquid mode. This can be effective for the ECMO patients, especially those with multiple organ failure.

本文言語English
ホスト出版物のタイトルProceedings of the Annual Conference on Engineering in Medicine and Biology
Place of PublicationPiscataway, NJ, United States
出版社Publ by IEEE
ページ1557-1559
ページ数3
13
pt 4
ISBN(印刷版)0780302168
出版ステータスPublished - 1991
外部発表はい
イベントProceedings of the 13th Annual International Conference of the IEEE Engineering in Medicine and Biology Society - Orlando, FL, USA
継続期間: 1991 10月 311991 11月 3

Other

OtherProceedings of the 13th Annual International Conference of the IEEE Engineering in Medicine and Biology Society
CityOrlando, FL, USA
Period91/10/3191/11/3