T1 - Deployment of stent graft in an excessively higher position above the renal artery induces a flow channel to the aneurysm in chimney endovascular aortic aneurysm repair
T2 - an in vitro study
AU - Shukuzawa, Kota
AU - Akaoka, Taku
AU - Umezu, Mitsuo
AU - Ohki, Takao
AU - Iwasaki, Kiyotaka
N1 - Funding Information:
Acknowledgements This study was supported by the Subsidy Program for Development of International Standards for Evaluation of Innovative Medical Devices and Regenerative Medicine Products, from the Ministry of Health, Labour and Welfare, Japan. The Excluder® was provided by W.L. Gore and Associates.
Publisher Copyright:
© 2019, The Japanese Society for Artificial Organs.
PY - 2019/9/13
Y1 - 2019/9/13
N2 - We aimed to investigate the influences of the sealing length above the renal artery (RA) on gutter formation, non-apposed regions between the aortic wall, stent graft (SG), and chimney graft and incidence of flow channel to the aneurysm in chimney endovascular aortic aneurysm repair (Ch-EVAR) using a juxtarenal abdominal aortic aneurysm model. Neck diameter and length of the silicone model were 24 and 4 mm, respectively. In double Ch-EVAR configuration using Advanta V12, 12 combinations were tested three times with two sizes [28.5 (20%-oversize) and 31 (30%-oversize) mm] of Excluder SG, three sealing lengths above the RA (10, 20, and 30 mm), and two deployment positions (anatomical and cross-leg). Gutter area, non-apposed region, and flow channels to the aneurysm were analyzed using micro-computed tomography. Average gutter area and non-apposed region of 30%-oversize SG were significantly smaller than those of 20%-oversize SG (p = 0.05). Furthermore, the non-apposed region of 30%-oversize SG with a 30-mm sealing length was significantly larger than that of the other sealing lengths. For 20%-oversize SGs, flow channel to the aneurysm was observed, except for the anatomical deployment with the sealing length of 10 mm. For 30%-oversize SGs, flow channel was absent, except for the SG with a 30-mm sealing length in both deployment positions. These flow channels were frequently formed through a valley space, existing in the lower unibody above the two limbs. Our data indicated that the optimal sealing length should be chosen in consideration of the device design difference due to the device diameter in Ch-EVAR.
AB - We aimed to investigate the influences of the sealing length above the renal artery (RA) on gutter formation, non-apposed regions between the aortic wall, stent graft (SG), and chimney graft and incidence of flow channel to the aneurysm in chimney endovascular aortic aneurysm repair (Ch-EVAR) using a juxtarenal abdominal aortic aneurysm model. Neck diameter and length of the silicone model were 24 and 4 mm, respectively. In double Ch-EVAR configuration using Advanta V12, 12 combinations were tested three times with two sizes [28.5 (20%-oversize) and 31 (30%-oversize) mm] of Excluder SG, three sealing lengths above the RA (10, 20, and 30 mm), and two deployment positions (anatomical and cross-leg). Gutter area, non-apposed region, and flow channels to the aneurysm were analyzed using micro-computed tomography. Average gutter area and non-apposed region of 30%-oversize SG were significantly smaller than those of 20%-oversize SG (p = 0.05). Furthermore, the non-apposed region of 30%-oversize SG with a 30-mm sealing length was significantly larger than that of the other sealing lengths. For 20%-oversize SGs, flow channel to the aneurysm was observed, except for the anatomical deployment with the sealing length of 10 mm. For 30%-oversize SGs, flow channel was absent, except for the SG with a 30-mm sealing length in both deployment positions. These flow channels were frequently formed through a valley space, existing in the lower unibody above the two limbs. Our data indicated that the optimal sealing length should be chosen in consideration of the device design difference due to the device diameter in Ch-EVAR.
KW - Chimney technique
KW - Endovascular aortic repair
KW - Gutter
KW - In vitro
KW - Juxtarenal aortic aneurysm
UR - http://www.scopus.com/inward/record.url?scp=85060436742&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85060436742&partnerID=8YFLogxK
U2 - 10.1007/s10047-019-01090-x
DO - 10.1007/s10047-019-01090-x
M3 - Article
C2 - 30663032
AN - SCOPUS:85060436742
SN - 1434-7229
VL - 22
SP - 200
EP - 206
JO - Journal of Artificial Organs
JF - Journal of Artificial Organs
IS - 3
ER -