抄録
We proposed and tested a 1-dimentional sharing block detector using dual photo-multiplier tubes (PMT). The detector consists of scintillator blocks and dual PMTs. The dual PMTs are optically coupled on the scintillator blocks overlapping in one dimensionaily. With this configuration, the scintillator size of the block detector can be reduced to be half of the conventional block detector in one direction and Isotropic spatial resolution can be obtained in transaxial and axial directions using commercially available dual PMTs. First, we measured the position and energy response for minimum configuration; one GSO block and two dual PMTs. The size of a single GSO was 2.9mm × 2.9mm × 20mm. The scintillators were arranged in 8 × 8 matrix with multi-layer optical film inserted partly between scintillators to obtain the resolved position response. We could obtain good position and energy responses with this configuration. Next, we fabricated a 1-D sharing block detector with 5 GSO blocks and 6 dual PMTs. With this configuration, we also obtained reasonable position and energy responses. These results indicated that the proposed 1-D sharing Mock detector will be a promising solution for developing high resolution or low cost PET systems.
本文言語 | English |
---|---|
ページ(範囲) | 3366-3369 |
ページ数 | 4 |
ジャーナル | IEEE Nuclear Science Symposium Conference Record |
巻 | 6 |
出版ステータス | Published - 2004 |
外部発表 | はい |
イベント | 2004 Nuclear Science Symposium, Medical Imaging Conference, Symposium on Nuclear Power Systems and the 14th International Workshop on Room Temperature Semiconductor X- and Gamma- Ray Detectors - Rome, Italy 継続期間: 2004 10月 16 → 2004 10月 22 |
ASJC Scopus subject areas
- 放射線
- 核物理学および高エネルギー物理学