Development of CdTe pixel detector for synchrotron radiation experiments

T. Hirono*, H. Toyokawa, Y. Furukawa, M. Kawase, T. Ohata, S. Wu, H. Ikeda, G. Sato, T. Takahashi, S. Watanabe

*この研究の対応する著者

研究成果: Article査読

6 被引用数 (Scopus)

抄録

This study describes the development of a CdTe pixel detector for high-energy X-ray diffraction experiments at synchrotron radiation facilities. We have developed three prototypes. The third prototype detector (SP8-02B) was designed with a pixel size of 200μm×200μm and a matrix of 20×50 pixels. The sensor was bonded to an application-specific integrated circuit with a preamplifier, shaper, window-type discriminator, and a 20-bit counter for each pixel. The SP8-02B detector improved the uniformity between pixels in one chip in comparison with the previous prototypes. The long-term stability was measured at various temperatures. A moderate cooling achieved good stability that was not affected by the polarization effect.

本文言語English
ページ(範囲)64-67
ページ数4
ジャーナルNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
731
DOI
出版ステータスPublished - 2013

ASJC Scopus subject areas

  • 器械工学
  • 核物理学および高エネルギー物理学

フィンガープリント