Development of integrated calibration assistant tool for Pixel-PSPMT type gamma camera

Toshio Modegi*, S. Koyasu, H. Mashino, S. Yamamoto, H. Iida

*この研究の対応する著者

研究成果: Conference contribution

1 被引用数 (Scopus)

抄録

We are developing small animal PET/ SPECT scanner systems and high-definition small pixel-PSPMT type gamma cameras for them. These sensors are causing problems such as two-dimensional spatial nonlinearity, ununiformity and vagueness of pixel areas in scintillated images by unbalanced PMT gains or additional scattered photons’ influences. For overcoming these problems, we are providing look-up table type image correction units in the cameras. In order to determine appropriate image correction LUT data efficiently, we propose an assistant tool, which provides either automatic or interactive functions for designing correction tables. In this paper, we describe several features of our developed tool, and results of comparative experiments among four types of XYdimensional geometric correction methods supported by our tool.

本文言語English
ホスト出版物のタイトルIFMBE Proceedings
編集者Sun I. Kim, Tae Suk Suh
出版社Springer Verlag
ページ1704-1707
ページ数4
1
ISBN(印刷版)9783540368397
DOI
出版ステータスPublished - 2007
外部発表はい
イベント10th World Congress on Medical Physics and Biomedical Engineering, WC 2006 - Seoul, Korea, Republic of
継続期間: 2006 8月 272006 9月 1

出版物シリーズ

名前IFMBE Proceedings
番号1
14
ISSN(印刷版)1680-0737
ISSN(電子版)1433-9277

Other

Other10th World Congress on Medical Physics and Biomedical Engineering, WC 2006
国/地域Korea, Republic of
CitySeoul
Period06/8/2706/9/1

ASJC Scopus subject areas

  • バイオエンジニアリング
  • 生体医工学

フィンガープリント

「Development of integrated calibration assistant tool for Pixel-PSPMT type gamma camera」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル