Estimation of the kinetic-optimized stimulus intensity envelope for drop foot gait rehabilitation

S. Tanabe*, S. Kubota, N. Itoh, T. Kimura, Y. Muraoka, A. Shimizu, Y. Kanada

*この研究の対応する著者

研究成果: Review article査読

抄録

The purpose of present study is to estimate the optimal stimulus intensity envelope for drop foot rehabilitation based on a kinetic perspective. The voluntary and electric-stimulated elicited dorsiflexion torque responses of 11 healthy subjects were measured. During dorsiflexion, we recorded the tibialis anterior (TA) electromyography (EMG) or the stimulation intensity at four angles of the ankle joint. From these measurements, we derived two approximate equations that estimate dorsiflexion produced by either voluntary contraction or by electrical stimulation using a sigmoid function and a stepwise-regression analysis. We then tested the predictive capability of the model using Pearson correlation. Both equations indicated high correlation coefficients. Finally, we derived a relation between the TA EMG amplitude and stimulation intensity. From the obtained equation, we determined the optimal stimulus amplitude. We assume that the derived stimulus intensity envelope, calculated from EMG amplitude and angle of ankle joint, satisfies kinetic demand.

本文言語English
ページ(範囲)210-216
ページ数7
ジャーナルJournal of Medical Engineering and Technology
36
4
DOI
出版ステータスPublished - 2012 5月

ASJC Scopus subject areas

  • 生体医工学