Flow injection analysis of potassium using an all-solid-state potassium-selective electrode as a detector

Shinichi Komaba, Junko Arakawa, Michiko Seyama, Tetsuya Osaka*, Ikuo Satoh, Sadako Nakamura

*この研究の対応する著者

研究成果: Article査読

19 被引用数 (Scopus)

抄録

The concentration of potassium was determined by a combination of flow injection analysis (FIA) with an all-solid-state potassium sensor detection. The all-solid-state potassium-selective electrode possessing long-term potential stability was fabricated by coating an electroactive polypyrrole/poly(4-styrenesulfonate) film electrode with a plasticized poly(vinyl chloride) membrane containing valinomycin. The simple FIA system developed in this laboratory demonstrated sensitivity identical to that in the batch system and achieved considerably rapid assay (150 samples h -1). Analyses of soy sauce and control serum samples by this FIA system yielded results in good agreement with those obtained by conventional measurements.

本文言語English
ページ(範囲)1293-1297
ページ数5
ジャーナルTalanta
46
6
DOI
出版ステータスPublished - 1998 8月