Fracture of nickel-titanium superelastic alloy in sodium hypochlorite solution

Ken'ichi Yokoyama*, Kazuyuki Kaneko, Eiji Yabuta, Kenzo Asaoka, Jun'ichi Sakai

*この研究の対応する著者

    研究成果: Article査読

    27 被引用数 (Scopus)

    抄録

    Fracture of the Ni-Ti superelastic alloy for endodontic instruments such as files was investigated with a sustained tensile-loading test in sodium hypochlorite (NaOCl) solution of various concentrations. It was found that the time to fracture was reduced when the applied stress exceeded the critical stress for martensite transformation. When the applied stress was higher than the critical stress, the 0.3 mm diameter wires of the Ni-Ti superelastic alloy sometimes fractured within 60 min. From the results of observations of the fracture surface using a scanning electron microscope, it was revealed that the fracture of the Ni-Ti superelastic alloy is significantly influenced by corrosion when the applied stress was higher than the critical stress for martensite transformation. The results of the present study suggest that one of the causes of the fracture of Ni-Ti files during clinical use is corrosion under the applied stress above the critical stress for martensite transformation in NaOCl solution.

    本文言語English
    ページ(範囲)43-48
    ページ数6
    ジャーナルMaterials Science and Engineering A
    369
    1-2
    DOI
    出版ステータスPublished - 2004 3月 25

    ASJC Scopus subject areas

    • 材料科学(全般)

    フィンガープリント

    「Fracture of nickel-titanium superelastic alloy in sodium hypochlorite solution」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

    引用スタイル