抄録
The Gamma Ray Spectrometer (GRS) will be onboard the Japanese lunar polar orbiter SELENE, to be launched in 2004, to obtain the information of element abundance on the entire lunar surface. GRS consists of a high-purity Ge detector as a main detector, which is an n-type Ge crystal with a high detection efficiency of 60 % encapsulated in an aluminum canister. In addition, BGO and plastic scintillators are employed for an anti-coincidence shield and for suppression of Compton photons and energetic ion backgrounds. GRS will be the first gamma ray spectrometer employing a Ge detector for the lunar mission. GRS adopted a Stirling cycle refrigerator, whose cooling capacity is 2.0 W at 80 K. The mechanical vibration from the cryosystem to the Ge detector is effectively reduced by the dual opposed pistons in the compressor and the flexible thermo-coupling link between the two components. An excellent energy resolution of < 3 keV at 1.33 MeV has been achieved in the prototype- model of GRS.
本文言語 | English |
---|---|
ホスト出版物のタイトル | IEEE Nuclear Science Symposium and Medical Imaging Conference |
編集者 | D. Merelli, J. Surget, M. Ulma |
巻 | 1 |
出版ステータス | Published - 2000 |
イベント | 2000 IEEE Nuclear Science Symposium Conference Record - Lyon 継続期間: 2000 10月 15 → 2000 10月 20 |
Other
Other | 2000 IEEE Nuclear Science Symposium Conference Record |
---|---|
City | Lyon |
Period | 00/10/15 → 00/10/20 |
ASJC Scopus subject areas
- コンピュータ ビジョンおよびパターン認識
- 産業および生産工学