Improvement of the CdTe diode detectors using a guard-ring electrode

Kazuhiro Nakazawa*, Kousuke Oonuki, Takaaki Tanaka, Yoshihito Kobayashi, Ken'ichi Tamura, Takefumi Mitani, Goro Sato, Shin Watanabe, Tadayuki Takahashi, Ryoichi Ohno, Akihito Kitajima, Yoshikatsu Kuroda, Mitsunobu Onishi

*この研究の対応する著者

研究成果: Article査読

68 被引用数 (Scopus)

抄録

Recent results from the Schottky CdTe diode detectors employing a guard-ring (GR) electrode are reported. A cathode electrode, made of platinum, was separated into an active electrode(s) and a surrounding GR. Typical leakage current of a device with active area of 2 × 2 mm2 and 0.5 mm thickness surrounded by a GR, is 7 and 20 pA at a bias of 100 and 500 V, respectively, operated at 20°C. Spectral resolution of this device is 0.93 and 1.2 keV (FWHM) at 59.5 and 122 keV, respectively, operated at 20°C with a bias of 800 V. Detailed study of the characteristics of these devices working as a gamma-ray detector is presented.

本文言語English
ページ(範囲)1881-1885
ページ数5
ジャーナルIEEE Transactions on Nuclear Science
51
4 I
DOI
出版ステータスPublished - 2004 8月

ASJC Scopus subject areas

  • 電子工学および電気工学
  • 原子力エネルギーおよび原子力工学

フィンガープリント

「Improvement of the CdTe diode detectors using a guard-ring electrode」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル