OXYGEN-EXCHANGE REACTION BETWEEN ARTIFICIAL LUNG DEVICE: THE HEME EMBEDDED IN POLYMERIZED LIPO LIPOSOME AS AN ARTIFICIAL OXYGEN CARRIER.

Makoto Yuasa*, Etsuo Hasegawa, Hiroyuki Nishide, Eishun Tsuchida

*この研究の対応する著者

研究成果: Article査読

4 被引用数 (Scopus)

抄録

An oxygen-supplying medium (OSM), which was prepared by embedding a synthetic heme complex in a bilayer of polymerized lipid liposome (polylipid liposome/heme), could blind molecular oxygen reversibly under physiological conditions. The oxygen-exchange reaction of OSM with blood was examined by using a liquid/liquid artificial lung device. For example, deoxy-blood (p//S(O//2) equals 0 torr) was passed countercurrent to oxy-OSM (p//S(O//2) equals 154 torr) through hollow fibers of the device to provide oxy-blood (p//b(O//2) equals 57 torr). The OSM was physicochemically and mechanically stable under strong shear stress during the passage through the hollow fibers and acted as the oxygen mediator to blood.

本文言語English
ページ(範囲)661-668
ページ数8
ジャーナルJournal of macromolecular science. Chemistry
A24
6
出版ステータスPublished - 1987