Oxygen transfer performance in the membrane curved tube oxygenator

Kazuo Tanishita*, Kanji Hoshino

*この研究の対応する著者

研究成果: Article査読

抄録

Based on the evaluation of oxygen transfer performance in terms of non-dimensional variables such as oxygen Sherwood number, fractional saturation and non-dimensional length, the conditions achieving the high performance of gas exchange is discussed. The non-dimensional length at the rated condition, in which the oxygen saturation at the outlet becomes 97%, was determined in the relation between the fractional saturation and non-dimensional length. Simultaneously the oxygen transfer performance expressed by oxygen Sherwood number has been evaluated in relation to the variation of non-dimensional length. Finally the design parameters such as internal diameter, tube length and residence time are determined based on the rated condition.

本文言語English
ページ(範囲)2383-2389
ページ数7
ジャーナルNippon Kikai Gakkai Ronbunshu, B Hen/Transactions of the Japan Society of Mechanical Engineers, Part B
55
516
出版ステータスPublished - 1989 8月

ASJC Scopus subject areas

  • 機械工学

フィンガープリント

「Oxygen transfer performance in the membrane curved tube oxygenator」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル