Ripple reduction control of the undulation pump total artificial heart

Itsuro Saito*, Tsuneo Chinzei, Shuichi Mochizuki, Yusuke Abe, Takashi Isoyama, Kiyotaka Iwasaki, Takafumi Suzuki, Tatsuro Karita, Toshiya Ono, Akimasa Kouno, Mitsuhiko Ishimaru, Atsushi Baba, Toshinaga Ozeki, Takahiro Tohyama, Shin Ichi Kobayashi, Kou Imachi

*この研究の対応する著者

研究成果: Article査読

抄録

An undulation pump total artificial heart (UPTAH) in which the revolutions of the motor are converted to undulation motion of a disk has been developed. In an experiment, a goat using the UPTAH survived for 54 days. However, a large ripple was observed in the device's output pressure and flow waveform. In calculating the spectrum of the ripple, we found that the ripple mainly comprised 2 frequency sine waves: 1 having the same frequency as and 1 having double the frequency of the motor revolutions. To reduce the ripple, 2 sine waves, 1 having the same frequency as and 1 having double the frequency of the motor revolutions, were provided to the motor current to modulate the pulse width of the pulse width modulation controlling the motor revolutions. This ripple control method reduced the pressure ripple by 90% in a mock circulation and by 70% in animal experiments. These results revealed that the ripple generated in the UPTAH could be controlled through the use of motor control software.

本文言語English
ページ(範囲)40-44
ページ数5
ジャーナルArtificial Organs
26
1
DOI
出版ステータスPublished - 2002

ASJC Scopus subject areas

  • バイオエンジニアリング
  • 医学(その他)
  • 生体材料
  • 生体医工学

フィンガープリント

「Ripple reduction control of the undulation pump total artificial heart」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル