Y1 - 2014/11/1
N2 - Thrombogenic reactions under multiple interactions of pharmacological agents, doses, and materials have not been well understood yet. The aim of this study was to investigate the ability to simultaneously compare thrombogenic reactions to different combinations of anticoagulants, doses, and blood-contacting materials, in a single human blood using an in vitro test method. Four venous blood samples were drawn from each of six healthy volunteers into syringes that contained two different amounts of heparin and argatroban to set the activated clotting time (ACT) to approximately 200 or 500 s, respectively. The four blood samples from each volunteer were immediately poured into two clinical-grade extracorporeal circulation tubes: a polyvinyl chloride (PVC) tube and a poly(2-methoxyethyl acrylate)-coated (PMEA) PVC tube. These tubes with an inner diameter of 12.7 mm were rotated at 183 rpm in a 37
AB - Thrombogenic reactions under multiple interactions of pharmacological agents, doses, and materials have not been well understood yet. The aim of this study was to investigate the ability to simultaneously compare thrombogenic reactions to different combinations of anticoagulants, doses, and blood-contacting materials, in a single human blood using an in vitro test method. Four venous blood samples were drawn from each of six healthy volunteers into syringes that contained two different amounts of heparin and argatroban to set the activated clotting time (ACT) to approximately 200 or 500 s, respectively. The four blood samples from each volunteer were immediately poured into two clinical-grade extracorporeal circulation tubes: a polyvinyl chloride (PVC) tube and a poly(2-methoxyethyl acrylate)-coated (PMEA) PVC tube. These tubes with an inner diameter of 12.7 mm were rotated at 183 rpm in a 37
KW - Activated clotting time
KW - Anticoagulant
KW - In vitro thrombogenicity test
KW - Platelet activation
KW - Total plasma protein adsorption
UR - http://www.scopus.com/inward/record.url?scp=84908118186&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84908118186&partnerID=8YFLogxK
U2 - 10.1002/jbm.b.33145
DO - 10.1002/jbm.b.33145
M3 - Article
C2 - 24652689
AN - SCOPUS:84908118186
SN - 1552-4973
VL - 102
SP - 1605
EP - 1612
JO - Journal of Biomedical Materials Research - Part B Applied Biomaterials
JF - Journal of Biomedical Materials Research - Part B Applied Biomaterials
IS - 8
ER -