Time over Threshold based digital animal PET (TODPET)

K. Shimazoe*, Y. Wang, H. Takahashi, K. Kamada, M. Yoshino, J. Kataoka, Y. Yamaya, T. Yanagida, A. Yoshikawa, K. Kumagai

*この研究の対応する著者

研究成果: Conference contribution

15 被引用数 (Scopus)

抄録

We have developed Time over Threshold (ToT) based Pr:LuAG-APD PET (TODPET) tomograph with a mixed signal front-end. The tomograph consists of 8 block detectors, each of which is composed of a 12 x 12 array of 2 x 2 x 10mm 3 Pr:LuAG crystals individually coupled with 12 x 12 UV-enhanced APD arrays. The APDs are individually read out with a custom-designed Time over Threshold ASIC and FPGA readout system. Developed PET tomograph has the energy resolution of 10% and the time resolution of 4.2ns. The 1.76mm spatial resolution (FWHM) is achieved for the first result.