Use of puncture force measurement to investigate the conditions of blood vessel needle insertion

Yo Kobayashi*, Ryutaro Hamano, Hiroki Watanabe, Jaesung Hong, Kazutaka Toyoda, Makoto Hashizume, Masakatsu G. Fujie

*この研究の対応する著者

研究成果: Article査読

13 被引用数 (Scopus)

抄録

Central venous catheterization involves venous puncture and catheter insertion for transfusions. Quantitative conditions that facilitate insertion of the needle, such as the angle and velocity of insertion, have not been clarified. We previously developed a robotic system for guiding the needle along a specified puncture path with high precision and are currently implementing a hardware design for a robotic system to assist in blood vessel puncture. In this study, we proposed the insertion angle and velocity for stopping the needle in a blood vessel, assuming that a robotic system such as ours is used during the procedure. We inserted a needle into a segment of porcine jugular vein and obtained the puncture reaction force. Evaluation indices were the magnitude of the sudden decrease in reaction force at the point at which the needle advances and the length of time that the needle is present within the vein. Results indicated that the conditions under which it was easiest to stop the needle inside the vein were an insertion angle range of 10-20 and an insertion velocity of 3 mm/s.

本文言語English
ページ(範囲)684-689
ページ数6
ジャーナルMedical Engineering and Physics
35
5
DOI
出版ステータスPublished - 2013 5月

ASJC Scopus subject areas

  • 生体医工学
  • 生物理学

フィンガープリント

「Use of puncture force measurement to investigate the conditions of blood vessel needle insertion」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル