TY - GEN
T1 - Validation of accuracy of liver model with temperature-dependent thermal conductivity by comparing the simulation and in vitro RF ablation experiment
AU - Watanabe, Hiroki
AU - Yamazaki, Nozomu
AU - Isobe, Yosuke
AU - Lu, Xiaowei
AU - Kobayashi, Yo
AU - Miyashita, Tomoyuki
AU - Ohdaira, Takeshi
AU - Hashizume, Makoto
AU - Fujie, Masakatsu G.
PY - 2012/12/14
Y1 - 2012/12/14
N2 - Radiofrequency (RF) ablation is increasingly used to treat cancer because it is minimally invasive. However, it is difficult for operators to control precisely the formation of coagulation zones because of the inadequacies of imaging modalities. To overcome this limitation, we previously proposed a model-based robotic ablation system that can create the required size and shape of coagulation zone based on the dimensions of the tumor. At the heart of such a robotic system is a precise temperature distribution simulator for RF ablation. In this article, we evaluated the simulation accuracy of two numerical simulation liver models, one using a constant thermal conductivity value and the other using temperature-dependent thermal conductivity values, compared with temperatures obtained using in vitro experiments. The liver model that reflected the temperature dependence of thermal conductivity did not result in a large increase of simulation accuracy compared with the temperature-independent model in the temperature range achieved during clinical RF ablation.
AB - Radiofrequency (RF) ablation is increasingly used to treat cancer because it is minimally invasive. However, it is difficult for operators to control precisely the formation of coagulation zones because of the inadequacies of imaging modalities. To overcome this limitation, we previously proposed a model-based robotic ablation system that can create the required size and shape of coagulation zone based on the dimensions of the tumor. At the heart of such a robotic system is a precise temperature distribution simulator for RF ablation. In this article, we evaluated the simulation accuracy of two numerical simulation liver models, one using a constant thermal conductivity value and the other using temperature-dependent thermal conductivity values, compared with temperatures obtained using in vitro experiments. The liver model that reflected the temperature dependence of thermal conductivity did not result in a large increase of simulation accuracy compared with the temperature-independent model in the temperature range achieved during clinical RF ablation.
UR - http://www.scopus.com/inward/record.url?scp=84870795305&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84870795305&partnerID=8YFLogxK
U2 - 10.1109/EMBC.2012.6347292
DO - 10.1109/EMBC.2012.6347292
M3 - Conference contribution
C2 - 23367227
AN - SCOPUS:84870795305
SN - 9781424441198
T3 - Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS
SP - 5712
EP - 5717
BT - 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBC 2012
T2 - 34th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, EMBS 2012
Y2 - 28 August 2012 through 1 September 2012
ER -