Very low-power consumption wireless ECG monitoring system using body as a signal transmission medium

Takashi Handa*, Shuichi Shoji, Shinichi Ike, Sunao Takeda, Tetsushi Sekiguchi

*この研究の対応する著者

研究成果: Paper査読

104 被引用数 (Scopus)

抄録

A very low-power consumption wireless system for monitoring ECG (ElectroCardioGram) is proposed. It consists of ECG detector/transmitter located on the chest and the relay transmitter placed at the wrist. Between the detector part and relay transmitter, the signal is send as the AC micro current flown through the tissue of the body. Since this method obtains very low-power signal transmission of about 8μW(400mVRMS × 20μARMS), the system can be driven by very small power source like a thin film battery. The similar wireless transmission method can be applied for other medical signal/data sensing micro systems sharing the same relay transmitter.

本文言語English
ページ1003-1006
ページ数4
出版ステータスPublished - 1997 1月 1
イベントProceedings of the 1997 International Conference on Solid-State Sensors and Actuators. Part 2 (of 2) - Chicago, IL, USA
継続期間: 1997 6月 161997 6月 19

Other

OtherProceedings of the 1997 International Conference on Solid-State Sensors and Actuators. Part 2 (of 2)
CityChicago, IL, USA
Period97/6/1697/6/19

ASJC Scopus subject areas

  • 工学(全般)

フィンガープリント

「Very low-power consumption wireless ECG monitoring system using body as a signal transmission medium」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル